»ó¼¼ ±Ô°Ý
»ó¼¼ ±Ô°Ý Ç¥
Cat. No Model ¿ë·® ¼öºÐÀ² ÃÖ¼ÒÇ¥½Ã
M11-640-101 MB120 120g 0.01%
M11-640-111 MB200 200g 0.01%
M11-640-121 MB54 50g 0.001%
»ó¼¼ ±Ô°Ý Ç¥
Cat. No M11-640-101 M11-640-111 M11-640-121
Model MB 120 MB 200 MB 54
Measurement Value
Readability 0.01% 0.01% 0.00%
Repeatability (sd) for initial sample wt.= 2g 0.10% 0.10% 0.05%
Repeatability (sd) for initial sample wt.= 10g 0.03% 0.03% 0.02%
Evaluation
Units of Measurement % M, %D, %R (ratio), Gm/Lt, %M Atro, %D Atro
Data Storage Last 1000 Moisture Determination results
Weighing Balance
Max. Sample weight 120 g 200 g 50 g
Minimum Sample Weight 0.2 g 0.2 g 0.2 g
Readability 0.001 g 0.001 g 0.1mg
Drying Unit
Technology Infra Red Halogen
Temperature range 30¨¬ C to 200¨¬ C
Temperature increment 1¨¬ C
Heating Profile Standard, Gentle, Rapid, Steps, High Temperature
User Support
Method Memory 40
Batch Memory 100
Method Designation Alpha Numeric
Serial Number YES
Display Graphical LCD display with LED backlit
Data Interface RS232C Bidirectional, USB, Ethernet, PS2
Print GLP / GMP Compliant, User Configurable
Miscellaneous
Power Supply 230V / 50Hz or 115V / 60Hz
Power Consumption (Watt) Max 415
Accessory
Cat. No Model Description
B02-640-330 CPR 02 Data Printer